| Flowmeter
Series > KEM > C-FLOW > KCE 4000 Transmitter |
| |
| Introduction KCM
Transducer KCE
4000 Transmitter |
| |
| P.1 P.2 |
| |
| KCE
4000 Transmitter |
| |
| General |
|
| Housing
(WG): |
- plastic
for wall mounting with transparent cover
-
200 x 215 x 168 (B x H x D) without cable glands
- weight:
approx. 1.5 kg
- protection
class: IP 54/DIN EN 60529
|
| Housing
(SG): |
- plastic
for panel mounting as per DIN 43700/IEC 61554
- optionally
with transparent cover
- 192
x 96 x 205 (B x H x D) without cover
- 196
x 99 x 241 (B x H x D) with cover
- panel
board cut out: 186 x 92
- weight:
approx. 1 kg
- protection
class: IP 42 (IP 52 with cover)
|
| Programming: |
|
| Display: |
- illuminated
two-line alpha numerical LCD display for
- measuring
values and parameter setting,
- 2x
20 characters, character size 2.4 x 4.7 mm
|
| Interface: |
|
| Temperature: |
-
storage and transport: -25 up to +75 ¢XC
- operation:
0 up to 50 ¢XC
|
| EMC: |
- according
to EN 50 081-2 and EN 50 082-2
|
| Supply
voltage: |
|
| Power
consumption: |
|
Ex-Protection:
|
II
(2) G [EEx ib] IIC
|
|
| |
| Analogue
Outputs |
|
| Voltage
output: |
- 2
off 0-5 V
- resolution:
12 bit
- linearity:
¡Ó0.05% of final value
- temperature
drift: 0.05 % per 10 K
- load:
> 10 k
- scaled
output of flow rate
|
| Current
output: |
- 1
off 0/4-20 mA, active, galvanically free
- resolution:
12 bit
- linearity:
¡Ó0.05% of final value
- temperature
drift: 0.05% per 10 K
- load:
< 800 £[
- scaled
output of flow rate or job total, density or
temperature
|
|
| |
| Pulse
Outputs |
|
| Frequency
range: |
|
| Output
signal: |
- adjustable
via jumpers
- open
collectior: UCE < 30 V, ICE<
50mA
- push
pull: Imax 20 mA
- output
of flow rate
|
|
| |
| Switch
Outputs (2 off) |
|
| Output
signal: |
- adjustable
via jumpers
- open
collectior: UCE < 30 V, ICE<
50mA
- push
pull: Imax 20 mA
|
| Use: |
- limit
1 for job total, fault or flow rate
-
limit 2 for job total, cycle or flow rate
|
|
| |
| Switch
Outputs (2 off) |
|
| Output
signal: |
- passive
"on" > 4 V, "off" <
1 V
|
| Use: |
- 1x
for job total reset
- for
offset
|
|
| |
|
Marking of the Transducer KCM |
| Type
KCM ********** |
Ui
= 10.6 V |
| S/N
************** |
Ii
= 80 mA |
II
2G EEx ib IIC T2-T4 |
Pi
= 0.22 W |
| Tamb.
-30 up to + 50 ¢XC |
Li
= 3 mH |
| Tmed.max.
+125¢XC |
|
|
| |
| Important
Note |
| The
transducer KCM **-Ex must only be operated with the associated
evaluation electronics transmitter KCE **-Ex. |
| |
| Marking
of the Transducer KCE |
| Type
KCM ********** |
Uo
= 10.6 V |
| S/N
************ |
Io
= 80 mA |
II
2G [ EEx ib] IIC |
Po
= 0.22 W |
| Tamb.
0 up to + 50 ¢XC |
Lo
= 4 mH |
| Ub
+24V/DC ¡Ó15% |
Co
= 2.3 £gF |
| |
Um
= 253 V/50 Hz |
|
| Important
Note |
The
transmitter KCE*** must not be operated within hazardous
areas.
KCM**-Ex and KCE*** must be connected with the ready-to-wire
cable by KEM (L=0.45 mH/km, C = 160 nF/km). The max. cable
length has to be below 75 metres and max. inductivities
and capacities for the respective zone have to be considered. |
| |