| Deviation
flow measurement: |
Less
than ¡Ó0.3% of max. flow. |
| Deviation
current output: |
Cf.
fig. 3, plus ¡Ó0.3% of range for current output. |
| Linearity: |
See
fig. 3. |
| Repeatability: |
Better
than 0.5 x deviation, see fig. 3. |
| Ambient
temperature effect: |
Less
than 0.04%/10¢XC. |
| Voltage
supply effect: |
Less
than 0.01%/10%. |
| Max.
capacity: |
| PD
340 - |
C
25: 8 m3/h. |
| |
C
38: 20 m3/h. |
| |
C
51: 40 m3/h. |
| |
C
63: 80 m3/h. |
|
| Measurement
of temperature: |
| Deviation: |
Less
than ¡Ó0.9¢XC with Pt-100 detector Din 43760.
|
| Limits: |
From
- 30¢XC to +100¢XC. |
|
| Ambient
temperature: |
From
-10¢XC to + 50¢XC. |
| Power
supply: |
24
V AC ¡Ó 15%, 50/60 c/s or 24 V DC ¡Ó 15%. |
| Power
consumption: |
Max.
6 W. |
| Liquid: |
| Conductivity: |
Min.
5mS/cm. |
| Temperature: |
From
- 30¢XC to +100¢XC. |
| Pressure: |
Max.
10 bar. |
|
| Outputs: |
The
transmitter can give two different output signals,
pulse signal or current signal. External voltage
supplies are required for both types of signals. |
| Pulse
signals: |
 |
| Current
signal: |
 |
| Material: |
| Electrodes: |
Stainless
steel AISI 316. |
| Metering
pipe: |
Stainless
steel AISI 316. |
| Coating
inside metering pipe: |
FEP
Teflon |
| Housing: |
PPO
Noryle |
|
| Connections: |
Clamp
pipe coupling DS/ISO 2852. |